Sign In

  • Forgot your password?
  • Need a new account?

Register


STM3013

β Detector, Input Voltage +11.5~ + 12.5V, Max Input Current ≤60mA(+12V); ≤10mA(-12V), Effective Area Φ25mm, Output Pulse Llogic TTL, Output Pulse Amplitude 5±0.2V, Output Pulse Width 500±100ns, Background Counting Rate Max. 3 s^-1, Observed Counting Rate 26.5-28k s^-1, 1 Hour Instability Typ. 0.2%

MODEL:STM3013

6-8 weeks Drawings

Save To My List

* The prices are indicative only. Please contact us for an official quote. This is single product page, for detail specification please refer to product main page


Compare Model Drawings & Specs Availability Reference Price
STM1111
β Detector, Input Voltage +11.5~ + 12.5V, Max Input Current 70mA, Effective Area Φ10mm, Output Pulse Logic +TTL, Output Pulse Amplitude 5±0.5V, Output Pulse Width 500±100ns, Background Counting Rate Max. 3 s^-1, Observed Counting Rate 8.8-9k s^-1, 1 Hour Instability Typ. 0.2%
6-8 weeks Request for quote

Compatible

Compare Model Drawings & Specs Availability Reference Price

Product Specifications

 Paramater Min.  Typ. Max. Unit
 Input Voltage ±11.5~±12.5 V
 Max Input Current ≤60(+12V); ≤10(-12V) mA
 Effective Area Φ25 mm
 Output Pulse Llogic TTL /
 Output Pulse Amplitude 5±0.2 V
 Output Pulse Width 500±100 ns
 Background Counting Rate / / 3 s-1
 Observed Counting Rate(1) 26.5-28k s-1
 1 Hour Instability(2) / 0.2 0.4 %
Work Environment Temperature   +5~+40
Humidity Max. 90%RH /
Storage Environment Temperature   -20~+50
Humidity Max. 93%RH /
 Test Ambient Temperature 25

 Note:

(1)Test with a 60μCi 14C source 3mm from the detector.

(2)1-hour instability test process: The count rate of detector acquisition was 6.0±0.5ks, the collection time of each data was set to 4min, and 15 data were collected cumulatively (1 hour) for each set of data. The absolute value of deviation between each value of each set of data and the mean value was calculated. The three groups were tested continuously (3 hours), and the maximum value of the three groups was taken as the 1-hour operating instability test value. 

STM3013 - Parameter

STM1111 - Parameter

STM3013 - Download

STM1111 - Download